A New Thermal Diffusion Flow Probe for Continuous Monitoring of Cerebral Blood Flow in Small Animals

The Peltier stack, which allows quantitative measurement of cerebral blood flow (CBF) by means of thermal diffusion, has a probe too large for use in small animals. However, it is difficult to measure CBF by a thermal diffusion method involving the use of a constantan heating wire, because of heat c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Neurologia medico-chirurgica 1989, Vol.29(12), pp.1078-1084
Hauptverfasser: KUWAYAMA, Naoya, FUKUDA, Osamu, KOSHU, Keiji, ENDO, Shunro, SAITO, Tateo, TAKAKU, Akira
Format: Artikel
Sprache:eng ; jpn
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Peltier stack, which allows quantitative measurement of cerebral blood flow (CBF) by means of thermal diffusion, has a probe too large for use in small animals. However, it is difficult to measure CBF by a thermal diffusion method involving the use of a constantan heating wire, because of heat conduction between the two gold plates. The authors developed a new thermal diffusion flow probe, using a constantan wire as a heat source rather than the Peltier stack. With the new probe, separation of the gold plates and attachment of a pair of long thermocouples minimize heat conduction between the two plates. Moreover, the probe itself is considerably smaller than that of the Peltier stack. The new probe was inserted to the subdural space of rabbits and the voltage (V; mV) was measured with an amplifier by the circuit of constant current method. CBF (F; ml/100 g/min) was measured simultaneously by the hydrogen clearance method in the adjacent cortex. A regression equation of F=29, 111(1/V-1/226) was obtained between 35 pairs of F and V (r=0.92, p
ISSN:0470-8105
1349-8029
DOI:10.2176/nmc.29.1078