Traumatic fracture of flexible anterior chamber intraocular lenses

Two patients had fracture of a flexible, open-loop, anterior chamber intraocular lens (IOL) caused by ocular trauma. Case 1 was a 78-year-old woman with a three-pointfixation IOL. The two portions of the fractured lens were removed from the anterior chamber. Visual acuity at 10 months postoperativel...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of cataract and refractive surgery 1997-11, Vol.23 (9), p.1418-1420
Hauptverfasser: Capoferri, Carlo, Vacchi, Saulo, Tafi, Aldo, Voglini, Ida Maria
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two patients had fracture of a flexible, open-loop, anterior chamber intraocular lens (IOL) caused by ocular trauma. Case 1 was a 78-year-old woman with a three-pointfixation IOL. The two portions of the fractured lens were removed from the anterior chamber. Visual acuity at 10 months postoperatively was hand motion as a result of corneal decompensation. Case 2 was a 36-year-old man with a four-point-fixation IOL that broke and dislocated into the vitreous. A pars plana vitrectomy was performed; the two lens portions were removed and exchanged with a sutured posterior chamber lens. Best corrected visual acuity 4 months postoperatively was 20/20. In both cases, the lens fracture took place at the loop-optic junction and with minimal visible damage to the eye structures.
ISSN:0886-3350
1873-4502
DOI:10.1016/S0886-3350(97)80126-5