The Influence of High-Frequency Jet Ventilation With Varying Cardiac-Cycle Specific Synchronization on Cardiac Output in ARDS
Previous studies have shown “beat-to-beat” variation in systemic BP with high-frequency jet ventilation (HFJV). However, it is not clear if such changes are paralleled by changes in cardiac output. To characterize the effect of HFJV near or equal to the heart rate (HR) on beat-to-beat cardiac output...
Gespeichert in:
Veröffentlicht in: | Chest 1997-12, Vol.112 (6), p.1600-1606 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Previous studies have shown “beat-to-beat” variation in systemic BP with high-frequency jet ventilation (HFJV). However, it is not clear if such changes are paralleled by changes in cardiac output.
To characterize the effect of HFJV near or equal to the heart rate (HR) on beat-to-beat cardiac output in an adult human subject with ARDS.
Case study.
ICU, university teaching hospital.
One patient with end-stage liver disease complicated by sepsis, severe pancreatitis, ARDS, and multisystem organ failure.
The patient was intubated, sedated, paralyzed, and ventilated with controlled mechanical ventilation (CMV). Ventilatory mode was then switched to HFJV at fixed frequencies (f) near but not equal to the HR (f=100, 110, and 120 beats/min; HR=108/min). HFJV was then synchronized to the ECG such that f and HR were equal. Continuous cardiac output (COc) was monitored during change of ventilator mode from CMV to fixed-rate HFJV to synchronized HFJV, then followed through progressive delays in jet triggering within the cardiac cycle during the synchronous HFJV mode. COc was monitored by arterial pulse-contour analysis, allowing assessment of beat-to-beat changes in cardiac output.
A cyclic variation in COc equal to the beat frequency difference between f and HR was observed (harmonic interaction) during fixed-rate HFJV. This COc oscillation was abolished during synchronous HFJV. COc was significantly greater during systolic synchronous HFJV as compared to diastolic synchronous HFJV or fixed-rate HFJV (10.1 to 9.0 [p |
---|---|