Development of a Low Flow Resistance Intravenous Oxygenator
A potentially attractive support device for patients with acute respiratory failure is an intravenous membrane oxygenator. One problem, however, is that the membrane surface area required for sufficient gas exchange can unduly increase vena cavai pressure drop and impede venous return. The purpose o...
Gespeichert in:
Veröffentlicht in: | ASAIO journal (1992) 1997-09, Vol.43 (5), p.M731-M731 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A potentially attractive support device for patients with acute respiratory failure is an intravenous membrane oxygenator. One problem, however, is that the membrane surface area required for sufficient gas exchange can unduly increase vena cavai pressure drop and impede venous return. The purpose of this study was to design and develop an intravenous oxygenator that would offer minimal venous flow resistance in situ. The device uses a constrained fiber bundle of smaller cross sectional size than the vena cava so as to effect an intentional shunt flow of venous blood around the fiber bundle and reduce the venous pressure drop caused by the device. A pulsating balloon within the fiber bundle redirects part of this shunt flow into reciprocating flow in and out of the fiber bundle. This offers dual advantages1) Blood flow through the fiber bundle is mainly perpendicular to the fibers; and 2) the requisite energy for driving flow comes largely |
---|---|