Transient Data Stream Dissociation in Computerized Data Acquisition System Masquerading as a Sensing Abnormality

During testing of a CPI model 1715 ICD, an apparent sensing abnormality was noted following shock delivery for VF. Close inspection of the recording prior to the defib‐rillation attempt revealed that the surface leads spontaneously lost 848 ms of data while the event marker was unaffected. Computer...