Transient Data Stream Dissociation in Computerized Data Acquisition System Masquerading as a Sensing Abnormality

During testing of a CPI model 1715 ICD, an apparent sensing abnormality was noted following shock delivery for VF. Close inspection of the recording prior to the defib‐rillation attempt revealed that the surface leads spontaneously lost 848 ms of data while the event marker was unaffected. Computer...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1997-10, Vol.20 (10), p.2500-2503
Hauptverfasser: NAIR, LAWRENCE A., GREENFIELD, RUTH ANN, YARGER, MICHAEL, NIBLEY, CARLETON, WHARTON, J. MARCUS
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:During testing of a CPI model 1715 ICD, an apparent sensing abnormality was noted following shock delivery for VF. Close inspection of the recording prior to the defib‐rillation attempt revealed that the surface leads spontaneously lost 848 ms of data while the event marker was unaffected. Computer simulations revealed that an inadequate buffer size for the ampHfied (surface ECC) data was the likely source of data loss. It is important to recognize tbat a discordance between surface leads and event marker may represent an abnormality in tlie data acquisition system and simulate an ICD or lead malfunction.