Upper Limit Ventricular Stimulation in Respiratory Rate Responsive Pacing due to Electrocautery
VAN HEMEL, N.M., et al.: Upper Limit Ventricular Stimulation in Respiratory Rate Responsive Pacing due to Electrocautery Transient programmed upper limit stimulation (150 bpm) was observed during repetitively utilized electrocautery in the beginning of an open‐heart surgical procedure in a patient w...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1989-11, Vol.12 (11), p.1720-1723 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | VAN HEMEL, N.M., et al.: Upper Limit Ventricular Stimulation in Respiratory Rate Responsive Pacing due to Electrocautery Transient programmed upper limit stimulation (150 bpm) was observed during repetitively utilized electrocautery in the beginning of an open‐heart surgical procedure in a patient with a minute ventilation rate responsive ventricuiar pacemaker. This tachycardia caused severe hemodynamic deterioration, and was also initiated by internal heart massage and manual ventilation. Considering the recommendations of the manufacturer, this series of serious events could have been prevented, when reprogramming to the inhibited mode had been executed in anticipation of the operation. |
---|---|