Long-term, totally implantable central venous access ports connected to a Groshong catheter for chemotherapy of solid tumours : Experience from 178 cases using a single type of device
The aim of this study was to examine the early and late complications rate of central venous access ports connected to the Groshong catheter for long-term chemotherapy delivering. All patients suffering from a neoplastic disease, who required long-term chemotherapy and underwent insertion of implant...
Gespeichert in:
Veröffentlicht in: | European journal of cancer (1990) 1997-07, Vol.33 (8), p.1190-1194 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to examine the early and late complications rate of central venous access ports connected to the Groshong catheter for long-term chemotherapy delivering. All patients suffering from a neoplastic disease, who required long-term chemotherapy and underwent insertion of implantable ports during a 21-month period (1 October 1994-30 June 1996) were prospectively studied. A single type of port was used, constructed of titanium and silicone rubber (Dome Port, Bard Inc., Salt Lake City, U.S.A), connected to an 8 F silastic Groshong catheter tubing (Bard Inc.). A team of different operators (two general surgeons, one interventional radiologist and four anaesthesiologists) was involved in inserting the port. All devices were placed in the operating room under fluoroscopic control. A central venous access form was filled in by the operator after the procedure and all ports were followed prospectively for device-related and overall complications. Data from the follow-up of these patients were entered in the form and collected in a database. Follow-up continued until the device was removed, the patient died or the study was closed. 178 devices, comprising a total of 32,089 days in situ, were placed in 175 patients. Three patients received a second device after removal of the first. Adequate follow-up was obtained in all cases (median 180 days, range 4-559). 138 devices (77.5%) were still in situ when the study was closed. Early complications included six pneumothoraxes, three arterial punctures and two revisions for port and/or catheter malfunction (overall early complications in 8 patients). Late complications included 3 cases (1.68% of devices) of catheter rupture and embolisation (0.093 episodes/1000 days of use), 2 cases (1.12% of devices) of venous thrombosis (0.062 episodes/1000 days of use), 1 case (0.56% of devices) of pocket infection (0.031 episodes/1000 days of use), and 4 cases (2.24% of devices) of port-related bacteraemias (0.124 episodes/1000 days of use). Infections were caused by coagulase-negative Staphylococcus aureus (4 cases) and Bacillus subtilis (1 case); they required port removal in 3 out of 5 cases. This study represents the largest published series of patients with totally implantable access ports connected to Groshong catheters; this device is a good option for long-term access to central veins and delivery of chemotherapeutic regimens, including continuous intravenous infusions. The low incidence of major complicati |
---|---|
ISSN: | 0959-8049 1879-0852 |
DOI: | 10.1016/S0959-8049(97)00039-7 |