Assessing validity of infant monitor alarms with event recording
We evaluated the performance of an event recorder system in a large, consecutive series of referred monitored patients to determine the relative incidence of true apnea and true bradycardia, false alarms, and alarms for movement or a loose lead. In addition, we developed an event classification syst...
Gespeichert in:
Veröffentlicht in: | The Journal of pediatrics 1989-11, Vol.115 (5), p.702-708 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated the performance of an event recorder system in a large, consecutive series of referred monitored patients to determine the relative incidence of true apnea and true bradycardia, false alarms, and alarms for movement or a loose lead. In addition, we developed an event classification system based on the reason for the event being recorded. The recorder stored transthoracic impedance and electrocardiogram signals on a floppy disk before, during, and after each monitor alarm. These events on 302 disks from 83 patients were analyzed and classified as true, false, or movement-loose lead. Of 14,131 events, only 8% were caused by apnea or bradycardia (true events). Of true events, 70% were triggered by apnea and 30% by bradycardia. These true events occurred in 48% of the patients. False alarms constituted 23%, and movement-loose lead 69%, of all events. Even when movement-loose lead events were excluded, nearly three of four events were found to be false. Event recording proved helpful clinically, allowing discontinuation of the monitor in 49% of patients, modification of monitor alarm settings, or reassurance and counseling for parents. |
---|---|
ISSN: | 0022-3476 1097-6833 |
DOI: | 10.1016/S0022-3476(89)80645-6 |