Diagnostic reliability of M-mode echocardiography for detecting mitral valve prolapse in 50 consecutive panic patients
Fifty consecutive panic patients had M-mode echocardiographs read independently by two cardiologists with expertise in echocardiography. In this prospective study, there was poor interrater reliability (22 of 50; K = 0.11) for diagnosis of mitral valve prolapse (MVP). On repeat evaluation 10 months...
Gespeichert in:
Veröffentlicht in: | Comprehensive psychiatry 1989-09, Vol.30 (5), p.369-375 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Fifty consecutive panic patients had M-mode echocardiographs read independently by two cardiologists with expertise in echocardiography. In this prospective study, there was poor interrater reliability (22 of 50; K = 0.11) for diagnosis of mitral valve prolapse (MVP). On repeat evaluation 10 months later there was also unacceptable intrarater reliability for each reader: 22 of 35 (K = 0.41) and 22 of 35 (K = 0.45). We conclude that M-mode echocardiography is clinically unreliable for establishing the diagnosis of mitral valve prolapse. These findings suggest that the variable reporting of M-mode-determined mitral valve prolapse in psychiatric populations may reflect differences among echocardiologists rather than differences in cardiac pathology. The clinical implications of these findings are discussed. |
---|---|
ISSN: | 0010-440X 1532-8384 |
DOI: | 10.1016/0010-440X(89)90002-3 |