Early femoral loosening in one design of cemented hip replacement
We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a th...
Gespeichert in:
Veröffentlicht in: | Journal of bone and joint surgery. British volume 1997-07, Vol.79 (4), p.603-608 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a thin cement mantle (p < 0.001) and excessive residual cancellous bone in the proximomedial region (p < 0.01). We recommend that the cement mantle around the prosthesis should be 2 to 3 mm and that further long-term studies are needed to evaluate the wear properties of titanium-nitride-coated titanium femoral heads. |
---|---|
ISSN: | 0301-620X 2049-4394 2044-5377 2049-4408 |
DOI: | 10.1302/0301-620x.79b4.0790603 |