Early femoral loosening in one design of cemented hip replacement

We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of bone and joint surgery. British volume 1997-07, Vol.79 (4), p.603-608
Hauptverfasser: MASSOUD, S. N, HUNTER, J. B, HOLDSWORTH, B. J, WALLACE, W. A, JULIUSSON, R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have studied aseptic loosening of the femoral component in 76 patients with primary total hip replacement using the Capital prosthesis. The mean follow-up was 26 months (10 to 37). Twelve femoral components (16%) were definitely and eight (10%) were possibly loose. They were characterised by a thin cement mantle (p < 0.001) and excessive residual cancellous bone in the proximomedial region (p < 0.01). We recommend that the cement mantle around the prosthesis should be 2 to 3 mm and that further long-term studies are needed to evaluate the wear properties of titanium-nitride-coated titanium femoral heads.
ISSN:0301-620X
2049-4394
2044-5377
2049-4408
DOI:10.1302/0301-620x.79b4.0790603