Compact Compliance Chamber Design for the Study of Cardiac Performance in Microgravity
A need was identified for a Mock Circulation System (MCS) of small size and weight that could function in a microgravity environment for the investigation of cardiovascular response to the weightlessness of space flight. Part of the MCS development involved the redesign of the compliance chamber fro...
Gespeichert in:
Veröffentlicht in: | ASAIO journal (1992) 1997-07, Vol.43 (4), p.316-320 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A need was identified for a Mock Circulation System (MCS) of small size and weight that could function in a microgravity environment for the investigation of cardiovascular response to the weightlessness of space flight. Part of the MCS development involved the redesign of the compliance chamber from a Penn State MCS using a coil spring instead of the leaf spring system employed in the Penn State system. The new compliance chambers achieve a weight reduction of 47% and a volume reduction of 64% over the original Penn State design. Testing showed the coil spring compliance chambers retained physiologic characteristics and adjustability by using coil springs of various stiffness, and functioned equivalently to the original Penn State design. |
---|---|
ISSN: | 1058-2916 1538-943X |
DOI: | 10.1097/00002480-199743040-00010 |