Predictors of fracture in the Accufix Atrial J lead
The Accufix atrial lead has a "J"-shaped retention wire at the distal end that has been reported to fracture. Our findings suggest that the more deformed the J, the higher the incidence of fracture.
Gespeichert in:
Veröffentlicht in: | The American journal of cardiology 1997-07, Vol.80 (2), p.229-231 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Accufix atrial lead has a "J"-shaped retention wire at the distal end that has been reported to fracture. Our findings suggest that the more deformed the J, the higher the incidence of fracture. |
---|---|