Variability and Reproducibility of Transcutaneous Oxygen Tension Measurements in the Assessment of Peripheral Vascular Disease

Transcutaneous oxygen tension (TcPO2) was measured according to a standard protocol in 43 limbs from 23 subjects, by use of oxygen-sensing electrodes attached to the chest (at a single site) or dorsum of the foot (at proximal and distal sites, located 1-2 cm apart). With the subjects supine, distal...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Angiology 1989-08, Vol.40 (8), p.695-700
Hauptverfasser: Rooke, Thom W., Osmundson, P.J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Transcutaneous oxygen tension (TcPO2) was measured according to a standard protocol in 43 limbs from 23 subjects, by use of oxygen-sensing electrodes attached to the chest (at a single site) or dorsum of the foot (at proximal and distal sites, located 1-2 cm apart). With the subjects supine, distal foot TcPO 2 differed from proximal by an average of ±7.4 mmHg, and in 80% of all limbs the differences were between -6 mmHg and + 15 mmHg. The two TcPO 2's were subsequently combined to yield a single "average" value for each foot. TcPO2 measurements were repeated twenty-four to forty-eight hours later; in the supine position the change in "average" foot TCPO 2 from the first to second measurement was ±6.9 mmHg, and in 80% of limbs the changes were between -11 mmHg and +9 mmHg. Similar reproducibility data were also obtained for ankle/brachial blood pressure indices (ABIs), chest TcPO2, and foot TcPO2 during three minutes of leg elevation at 30 degrees. The authors conclude that: (1) TcPO2 measurements from adjacent areas on the dorsum of the foot usually differ by 20-25% or less and (2) the short-term reproducibility of TcPO2 between studies is comparable to that for ABIS.
ISSN:0003-3197
1940-1574
DOI:10.1177/000331978904000802