Performance study of a miniature gamma ray scintillation vivo probe for tumor localization

We have developed a miniature gamma-ray endoscopic probe consisting of dual BGO detector probes for tumor detection inside the body cavities. The dual detector system was coupled with random coincidence to decrease the distant background radiation and to improve its spatial resolution for tumor loca...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of nuclear medicine 1997-05, Vol.11 (2), p.173-181
Hauptverfasser: Deloar, H M, Watabe, H, Hayashi, Y, Miyake, M, Nakamura, T, Takahashi, H, Yoshioka, T, Kanamaru, R, Fujiwara, T, Itoh, M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have developed a miniature gamma-ray endoscopic probe consisting of dual BGO detector probes for tumor detection inside the body cavities. The dual detector system was coupled with random coincidence to decrease the distant background radiation and to improve its spatial resolution for tumor localization. The performance of the probe was investigated with a point source and a water phantom. A solution of positron emitting 18F isotope was used as the source. Clinical trials of the probe were done to localize tumors on the skin surface of four subjects carrying tumors close to the body surfaces, into whom 67Ga-citrate and 18F-FDG radiopharmaceuticals were injected. Measurements indicated that the spatial resolution of the dual detector probes is around 1.5 times better than the single detector probe, and both single and dual detector endoscopic probe systems are capable of localizing a tumor on a large photon background. The endoscopic probe may be easier to insert inside body cavities due to the small crystal size and the flexible light guides. A single detector probe with higher sensitivity may be useful in searching for tumors over a wide intracavity area but a dual detector probe can be used for precise tumor localization. The detector probe may also be suitable for intraoperative observation.
ISSN:0914-7187
DOI:10.1007/BF03164830