A stopped-flow mixer device for a batch microcalorimeter application to NAD-NADase reaction
A new molded polypropylene, diamond-like carbon (DLC)-coated mixing cell has been developed for use in the batch microcalorimeter. Reagent volume can be varied from 25 μ1 to 100 μ1. A 10 microcalorie reaction heat can measure to 5%. Repeat reactions can be done as often as every 10 min for a fast re...
Gespeichert in:
Veröffentlicht in: | Journal of biochemical and biophysical methods 1989-03, Vol.18 (2), p.113-124 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new molded polypropylene, diamond-like carbon (DLC)-coated mixing cell has been developed for use in the batch microcalorimeter. Reagent volume can be varied from 25 μ1 to 100 μ1. A 10 microcalorie reaction heat can measure to 5%. Repeat reactions can be done as often as every 10 min for a fast reaction. Reactions can be started within 1 h or less after loading. A pre-equilibrator and a temperature-controlled syringe drive unit permit solutions to be stored at 4°C while being runt at any temperature from −20°C to 40°C. The kinetics and enthalpy of reaction of NAD-NADase have been measured.
ΔH is about 21 kcal/mol endothermic. |
---|---|
ISSN: | 0165-022X 1872-857X |
DOI: | 10.1016/0165-022X(89)90073-0 |