A stopped-flow mixer device for a batch microcalorimeter application to NAD-NADase reaction

A new molded polypropylene, diamond-like carbon (DLC)-coated mixing cell has been developed for use in the batch microcalorimeter. Reagent volume can be varied from 25 μ1 to 100 μ1. A 10 microcalorie reaction heat can measure to 5%. Repeat reactions can be done as often as every 10 min for a fast re...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biochemical and biophysical methods 1989-03, Vol.18 (2), p.113-124
Hauptverfasser: Berger, R.L., Mudd, C.P., Clem, T., Kolobow, T., Beile, E., Simons, P.C., Michel, S., McClintock, W.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new molded polypropylene, diamond-like carbon (DLC)-coated mixing cell has been developed for use in the batch microcalorimeter. Reagent volume can be varied from 25 μ1 to 100 μ1. A 10 microcalorie reaction heat can measure to 5%. Repeat reactions can be done as often as every 10 min for a fast reaction. Reactions can be started within 1 h or less after loading. A pre-equilibrator and a temperature-controlled syringe drive unit permit solutions to be stored at 4°C while being runt at any temperature from −20°C to 40°C. The kinetics and enthalpy of reaction of NAD-NADase have been measured. ΔH is about 21 kcal/mol endothermic.
ISSN:0165-022X
1872-857X