Use of an intraaortic balloon pump as a pneumatic ventricular assist device controller
Using a circulatory analogue, we investigated sequentially the performance of a dedicated ventricular assist device driver and an inttaaortic balloon pump when driving a pneumatic ventricular assist device. Each drive device was compared under identical pumping conditions at rates of 40 to 120 cycle...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 1989-05, Vol.47 (5), p.752-755 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using a circulatory analogue, we investigated sequentially the performance of a dedicated ventricular assist device driver and an inttaaortic balloon pump when driving a pneumatic ventricular assist device. Each drive device was compared under identical pumping conditions at rates of 40 to 120 cycles/min against two resistances. Our preliminary study showed that a modified intraaortic balloon pump could drive a pneumatic ventricular assist device as effectively as its dedicated driver. The necessary modifications to and possible further development of the intraaortic balloon in this role are discussed. |
---|---|