Using a circulatory analogue, we investigated sequentially the performance of a dedicated ventricular assist device driver and an inttaaortic balloon pump when driving a pneumatic ventricular assist device. Each drive device was compared under identical pumping conditions at rates of 40 to 120 cycle...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 1989-05, Vol.47 (5), p.752-755
Hauptverfasser: Macrae, D.J., Glenville, B., McCarthy, T., Cooper, L., Guerreiro, D., Ross, D.N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Using a circulatory analogue, we investigated sequentially the performance of a dedicated ventricular assist device driver and an inttaaortic balloon pump when driving a pneumatic ventricular assist device. Each drive device was compared under identical pumping conditions at rates of 40 to 120 cycles/min against two resistances. Our preliminary study showed that a modified intraaortic balloon pump could drive a pneumatic ventricular assist device as effectively as its dedicated driver. The necessary modifications to and possible further development of the intraaortic balloon in this role are discussed.