Improved vessel visualization in MR angiography by nonlinear anisotropic filtering

This paper deals with a preprocessing technique of magnetic resonance angiography (MRA) images, applied before maximum‐intensity‐projection (MIP). The purpose was to recover small low‐intensity vessels, visible in individual slices, but lost in MIP images that usually have higher background level th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 1997-06, Vol.37 (6), p.914-919
Hauptverfasser: Orkisz, Maciej M., Bresson, Christine, Magnin, Isabelle E., Champin, Olivier, Douek, Philippe C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper deals with a preprocessing technique of magnetic resonance angiography (MRA) images, applied before maximum‐intensity‐projection (MIP). The purpose was to recover small low‐intensity vessels, visible in individual slices, but lost in MIP images that usually have higher background level than the individual slices. The authors have developed a nonlinear three‐dimensional spatial filtering technique (called HD filter) based on anisotropic smoothing. The filter first searches for the local orientation of the vessel. It then performs a nonlinear smoothing in the vessel's local direction so as to avoid blurring its boundaries. Noise level reduction, contrast enhancement, and improved small vessel visibility achieved by this filter are illustrated on dynamic contrast‐enhanced subtraction MRA images of the lower limbs.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.1910370617