An evaluation of the SCOTI device

The Sonomatic Confirmation of Tracheal Intubation (SCOTI, Penlon, UK) device was evaluated in the laboratory with cut and uncut tracheal tubes and in 50 adult patients who underwent tracheal and oesophageal intubation with uncut tracheal tubes. We could not configure the SCOTI reliably with cut trac...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 1997-05, Vol.52 (5), p.453-456
Hauptverfasser: Haridas, R. P., Chesshire, N. J., Rocke, D. A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Sonomatic Confirmation of Tracheal Intubation (SCOTI, Penlon, UK) device was evaluated in the laboratory with cut and uncut tracheal tubes and in 50 adult patients who underwent tracheal and oesophageal intubation with uncut tracheal tubes. We could not configure the SCOTI reliably with cut tracheal tubes. In the clinical study using uncut tubes the SCOTI identified all oesophageal intubations, but was only able to identify 84% (95% CI: 73.8–94.2%) of tracheal intubations. We recommend caution in using this device and further independent evaluation to determine its reliability.
ISSN:0003-2409
1365-2044
DOI:10.1111/j.1365-2044.1997.111-az0108.x