On the evaluation of data from flow-dialysis experiments

Flow dialysis can be used to measure (i) ligand binding to macromoloecules and (ii) the size of transmembrane ion gradients. Generally an approximate method is used to calculate the binding or gradient parameters from the raw data. Here we present a simple but exact method and evaluate the errors th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biochemical and biophysical methods 1989-02, Vol.18 (1), p.53-64
Hauptverfasser: Westerhoff, H.V., Wiechmann, A.H.C.A., van Dam, K., Hellingwerf, K.J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Flow dialysis can be used to measure (i) ligand binding to macromoloecules and (ii) the size of transmembrane ion gradients. Generally an approximate method is used to calculate the binding or gradient parameters from the raw data. Here we present a simple but exact method and evaluate the errors that may arise when the approximate method is used to calculate the magnitude of ion gradients. In addition, equations are presented that allow for a correction for sampling from or additions to the upper compartment of a flow-dialysis vessel during the measurements. Setty and Hendler [(1982) J. Biochem. Biophys. Methods 7, 35–46] have reported artifacts in the measurement of ion-gradients caused by the addition of electron donors to the upper compartment of a flow-dialysis cell. Here we extend their observations and suggest additional methods to prevent such artifacts.
ISSN:0165-022X
1872-857X
DOI:10.1016/0165-022X(89)90013-4