Routine determination of radiochemical purity of 99mTc-MIBI

The manufacturer recommends that the radiochemical purity of 99mTc-MIBI be determined with alumina TLC plates developed in ethanol, but for routine use this is slow and requires care. After comparing a variety of systems, we obtained equivalent results using ITLC-SG strips developed with saline to d...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes Applied radiation and isotopes, 1989, Vol.40 (1), p.95-97
Hauptverfasser: Proulx, Annette, Ballinger, James R., Gulenchyn, Karen Y.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The manufacturer recommends that the radiochemical purity of 99mTc-MIBI be determined with alumina TLC plates developed in ethanol, but for routine use this is slow and requires care. After comparing a variety of systems, we obtained equivalent results using ITLC-SG strips developed with saline to determine [ 99mTc]pertechnetate and with acetone to determine 99mTc colloid; % 99mTc-MIBI was determined by difference from 100%. These separations are the opposite of those obtained with the routine polar 99mTc radiopharmaceuticals. Equivalent results were obtained even more rapidly by extracting 99mTc-MIBI from water into chloroform.
ISSN:0883-2889
DOI:10.1016/0883-2889(89)90182-2