Routine determination of radiochemical purity of 99mTc-MIBI
The manufacturer recommends that the radiochemical purity of 99mTc-MIBI be determined with alumina TLC plates developed in ethanol, but for routine use this is slow and requires care. After comparing a variety of systems, we obtained equivalent results using ITLC-SG strips developed with saline to d...
Gespeichert in:
Veröffentlicht in: | International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes Applied radiation and isotopes, 1989, Vol.40 (1), p.95-97 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The manufacturer recommends that the radiochemical purity of
99mTc-MIBI be determined with alumina TLC plates developed in ethanol, but for routine use this is slow and requires care. After comparing a variety of systems, we obtained equivalent results using ITLC-SG strips developed with saline to determine [
99mTc]pertechnetate and with acetone to determine
99mTc colloid; %
99mTc-MIBI was determined by difference from 100%. These separations are the opposite of those obtained with the routine polar
99mTc radiopharmaceuticals. Equivalent results were obtained even more rapidly by extracting
99mTc-MIBI from water into chloroform. |
---|---|
ISSN: | 0883-2889 |
DOI: | 10.1016/0883-2889(89)90182-2 |