End-tidal carbon dioxide (CO2) monitoring in small animals

A simple modification of a commercially available capnometer (Traverse Medical Monitors model 2000) is described which suits it for use with small animals such as rats. The modification involves the omission of the fluids separator from the gas pathway, while maintaining a connection to the separato...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brain research bulletin 1989-03, Vol.22 (3), p.579-582
Hauptverfasser: SMITH, K. J, FELTS, P. A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple modification of a commercially available capnometer (Traverse Medical Monitors model 2000) is described which suits it for use with small animals such as rats. The modification involves the omission of the fluids separator from the gas pathway, while maintaining a connection to the separator to permit the continued function of the "automatic zero" feature, which corrects for any zero-drift. The capnometer has several functions which recommend its use even when artificial ventilation is not required.
ISSN:0361-9230
1873-2747
DOI:10.1016/0361-9230(89)90116-0