Stability of radiofrequency magnetron sputtered calcium phosphate coatings under cyclically loaded conditions

The stability of radiofrequency (RF) magnetron sputtered calcium phosphate was studied under cyclically loaded conditions. The coatings were deposited on titanium bars and tested in either dry or wet conditions. X-ray diffraction (XRD), scanning electron microscopy (SEM), energy dispersive X-ray (ED...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomaterials 1997-03, Vol.18 (6), p.483-488
Hauptverfasser: Wolke, J.G.C., van der Waerden, J.P.C.M., de Groot, K., Jansen, J.A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The stability of radiofrequency (RF) magnetron sputtered calcium phosphate was studied under cyclically loaded conditions. The coatings were deposited on titanium bars and tested in either dry or wet conditions. X-ray diffraction (XRD), scanning electron microscopy (SEM), energy dispersive X-ray (EDX) analysis and Fourier transform infrared (FTIR) spectroscopy were used to characterize the assputtered and tested coatings. XRD demonstrated that the amorphous structure after annealing at 650 °C changed into a crystalline apatite structure. The residual stresses were determined by the XRD cos2ø method. These residual film stresses were influenced by the coating conditions and the crystalline sputtered coating showed the presence of compressive stresses. SEM demonstrated that, after cyclic loading conditions in air, the crystalline sputter-coated Ti-6AI-4V bars showed a partial coating loss. Furthermore, in wet conditions (simulated body fluid) only the heat-treated sputter-coated bars appeared to be stable. On the other hand, the amorphous coating only showed signs of delamination in the more highly stressed regions, while in the less stressed regions a Ca-P precipitate was formed. On the basis of these results we conclude that calcium phosphate coatings subjected to cyclic loading conditions show an important difference in fatigue behaviour when tested in either dry or wet conditions.
ISSN:0142-9612
1878-5905
DOI:10.1016/S0142-9612(96)00164-0