The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression
The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breat...
Gespeichert in:
Veröffentlicht in: | The European respiratory journal 1989-01, Vol.2 (1), p.71-77 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 77 |
---|---|
container_issue | 1 |
container_start_page | 71 |
container_title | The European respiratory journal |
container_volume | 2 |
creator | Verschakelen, JA Deschepper, K Clarysse, I Demedts, M |
description | The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved. |
doi_str_mv | 10.1183/09031936.93.02010071 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_78926000</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>78926000</sourcerecordid><originalsourceid>FETCH-LOGICAL-c405t-930224adc0bf309872da8466223c024ecc2a5cb7ead0f394ef6649a44498207b3</originalsourceid><addsrcrecordid>eNo9kE2LFDEQhoMo6-zoP1AIKN56rHxsd-coi7rCgpf1HNLpynSW9IdJWhkP_nYzTLunFFXP-wYeQt4wODDWio-gQDAl6oMSB-DAABr2jOyYUKoSAOI52Z2R6sy8JNcpPQKwWgp2Ra54A40EuSN_Hwak6BzaTGdHu4gmDzT5P0jN1NNlTnmNSOeJWhN8F032ZS5kLrmIafFlNccT9VO_2ux_IV0C5uGUxvkYzTLQ7kTHNWRf1jT4CU0suWOJptL0irxwJiR8vb178uPL54fbu-r--9dvt5_uKyvhJldKAOfS9BY6J0C1De9NK-uac2GBS7SWmxvbNWh6cEJJdHUtlZFSqpZD04k9-XDpXeL8c8WU9eiTxRDMhPOadNMqXkORtifyAto4pxTR6SX60cSTZqDP2vV_7VqVedNeYm-3_rUbsX8KbZ7L_f12N6mIdNFM1qcnrAHVqFK6J-8u2OCPw28fUafRhFBKmcb4yDXT5bN_MH2Ylg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>78926000</pqid></control><display><type>article</type><title>The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression</title><source>MEDLINE</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Verschakelen, JA ; Deschepper, K ; Clarysse, I ; Demedts, M</creator><creatorcontrib>Verschakelen, JA ; Deschepper, K ; Clarysse, I ; Demedts, M</creatorcontrib><description>The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved.</description><identifier>ISSN: 0903-1936</identifier><identifier>EISSN: 1399-3003</identifier><identifier>DOI: 10.1183/09031936.93.02010071</identifier><identifier>PMID: 2707404</identifier><language>eng</language><publisher>Leeds: Eur Respiratory Soc</publisher><subject>Abdomen - physiology ; Adult ; Biological and medical sciences ; Humans ; Investigative techniques of respiratory function ; Investigative techniques, diagnostic techniques (general aspects) ; Male ; Medical sciences ; Mouth - physiology ; Plethysmography - methods ; Posture ; Regression Analysis ; Respiration ; Ribs - physiology ; Vital Capacity</subject><ispartof>The European respiratory journal, 1989-01, Vol.2 (1), p.71-77</ispartof><rights>1989 INIST-CNRS</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c405t-930224adc0bf309872da8466223c024ecc2a5cb7ead0f394ef6649a44498207b3</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,777,781,4010,27904,27905,27906</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=7097919$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2707404$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Verschakelen, JA</creatorcontrib><creatorcontrib>Deschepper, K</creatorcontrib><creatorcontrib>Clarysse, I</creatorcontrib><creatorcontrib>Demedts, M</creatorcontrib><title>The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression</title><title>The European respiratory journal</title><addtitle>Eur Respir J</addtitle><description>The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved.</description><subject>Abdomen - physiology</subject><subject>Adult</subject><subject>Biological and medical sciences</subject><subject>Humans</subject><subject>Investigative techniques of respiratory function</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Mouth - physiology</subject><subject>Plethysmography - methods</subject><subject>Posture</subject><subject>Regression Analysis</subject><subject>Respiration</subject><subject>Ribs - physiology</subject><subject>Vital Capacity</subject><issn>0903-1936</issn><issn>1399-3003</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1989</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kE2LFDEQhoMo6-zoP1AIKN56rHxsd-coi7rCgpf1HNLpynSW9IdJWhkP_nYzTLunFFXP-wYeQt4wODDWio-gQDAl6oMSB-DAABr2jOyYUKoSAOI52Z2R6sy8JNcpPQKwWgp2Ra54A40EuSN_Hwak6BzaTGdHu4gmDzT5P0jN1NNlTnmNSOeJWhN8F032ZS5kLrmIafFlNccT9VO_2ux_IV0C5uGUxvkYzTLQ7kTHNWRf1jT4CU0suWOJptL0irxwJiR8vb178uPL54fbu-r--9dvt5_uKyvhJldKAOfS9BY6J0C1De9NK-uac2GBS7SWmxvbNWh6cEJJdHUtlZFSqpZD04k9-XDpXeL8c8WU9eiTxRDMhPOadNMqXkORtifyAto4pxTR6SX60cSTZqDP2vV_7VqVedNeYm-3_rUbsX8KbZ7L_f12N6mIdNFM1qcnrAHVqFK6J-8u2OCPw28fUafRhFBKmcb4yDXT5bN_MH2Ylg</recordid><startdate>19890101</startdate><enddate>19890101</enddate><creator>Verschakelen, JA</creator><creator>Deschepper, K</creator><creator>Clarysse, I</creator><creator>Demedts, M</creator><general>Eur Respiratory Soc</general><general>Maney</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19890101</creationdate><title>The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression</title><author>Verschakelen, JA ; Deschepper, K ; Clarysse, I ; Demedts, M</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c405t-930224adc0bf309872da8466223c024ecc2a5cb7ead0f394ef6649a44498207b3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1989</creationdate><topic>Abdomen - physiology</topic><topic>Adult</topic><topic>Biological and medical sciences</topic><topic>Humans</topic><topic>Investigative techniques of respiratory function</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Mouth - physiology</topic><topic>Plethysmography - methods</topic><topic>Posture</topic><topic>Regression Analysis</topic><topic>Respiration</topic><topic>Ribs - physiology</topic><topic>Vital Capacity</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Verschakelen, JA</creatorcontrib><creatorcontrib>Deschepper, K</creatorcontrib><creatorcontrib>Clarysse, I</creatorcontrib><creatorcontrib>Demedts, M</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The European respiratory journal</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Verschakelen, JA</au><au>Deschepper, K</au><au>Clarysse, I</au><au>Demedts, M</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression</atitle><jtitle>The European respiratory journal</jtitle><addtitle>Eur Respir J</addtitle><date>1989-01-01</date><risdate>1989</risdate><volume>2</volume><issue>1</issue><spage>71</spage><epage>77</epage><pages>71-77</pages><issn>0903-1936</issn><eissn>1399-3003</eissn><abstract>The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved.</abstract><cop>Leeds</cop><pub>Eur Respiratory Soc</pub><pmid>2707404</pmid><doi>10.1183/09031936.93.02010071</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0903-1936 |
ispartof | The European respiratory journal, 1989-01, Vol.2 (1), p.71-77 |
issn | 0903-1936 1399-3003 |
language | eng |
recordid | cdi_proquest_miscellaneous_78926000 |
source | MEDLINE; EZB-FREE-00999 freely available EZB journals |
subjects | Abdomen - physiology Adult Biological and medical sciences Humans Investigative techniques of respiratory function Investigative techniques, diagnostic techniques (general aspects) Male Medical sciences Mouth - physiology Plethysmography - methods Posture Regression Analysis Respiration Ribs - physiology Vital Capacity |
title | The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-18T00%3A58%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20effect%20of%20breath%20size%20and%20posture%20on%20calibration%20of%20the%20respiratory%20inductive%20plethysmograph%20by%20multiple%20linear%20regression&rft.jtitle=The%20European%20respiratory%20journal&rft.au=Verschakelen,%20JA&rft.date=1989-01-01&rft.volume=2&rft.issue=1&rft.spage=71&rft.epage=77&rft.pages=71-77&rft.issn=0903-1936&rft.eissn=1399-3003&rft_id=info:doi/10.1183/09031936.93.02010071&rft_dat=%3Cproquest_cross%3E78926000%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=78926000&rft_id=info:pmid/2707404&rfr_iscdi=true |