The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression
The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breat...
Gespeichert in:
Veröffentlicht in: | The European respiratory journal 1989-01, Vol.2 (1), p.71-77 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved. |
---|---|
ISSN: | 0903-1936 1399-3003 |
DOI: | 10.1183/09031936.93.02010071 |