Induced membrane hypo/hyper-mechanosensitivity: a limitation of patch-clamp recording
Practical limitations of the patch-clamp technique when recording mechanogated membrane ion channels are considered. Mechanical overstimulation of the patch or the cell from excessive suction/pressure protocols induces morphological and functional changes. In particular, the plasma membrane becomes...
Gespeichert in:
Veröffentlicht in: | Annual review of physiology 1997-01, Vol.59 (1), p.621-631 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Practical limitations of the patch-clamp technique when recording mechanogated membrane ion channels are considered. Mechanical overstimulation of the patch or the cell from excessive suction/pressure protocols induces morphological and functional changes. In particular, the plasma membrane becomes decoupled from the underlying cytoskeleton to form either membrane blebs (cell-attached) or ghosts (whole cell). As a consequence, a membrane ion channel may show either a decrease or an increase in its native mechanosensitivity or even acquire mechanosensitivity. The effect varies with ion channel and cell type and presumably arises because of a disruption of membrane-cytoskeleton interactions. We consider that such disruptions are a pathological consequence of excessive mechanical stress, either during or after seal formation, rather than an immutable consequence of patch-clamp recording. By careful attention to the suction/pressure protocols during sealing and throughout recording, such artifacts can be avoided. |
---|---|
ISSN: | 0066-4278 1545-1585 |
DOI: | 10.1146/annurev.physiol.59.1.621 |