The optimum dose of gamma radiation-heavy doses to low wear polyethylene in total hip prostheses

Wear volume, surface area and coefficient of friction of UHMWPE cup crosslinked with gamma radiation of 0, 50, 75, 100, 125, 150 and 200 Mrad sliding against an alumina ball were measured using a sphere-on-flat reciprocating type tribology testing machine. The effects of gamma radiation were scarcel...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of materials science. Materials in medicine 1997-01, Vol.8 (1), p.11-18
Hauptverfasser: Oonishi, H, Kuno, M, Tsuji, E, Fujisawa, A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Wear volume, surface area and coefficient of friction of UHMWPE cup crosslinked with gamma radiation of 0, 50, 75, 100, 125, 150 and 200 Mrad sliding against an alumina ball were measured using a sphere-on-flat reciprocating type tribology testing machine. The effects of gamma radiation were scarcely observed in coefficient of friction. The coefficient of friction under lubricated (distilled water) and non-lubricated testing conditions was 0.08 to 0.12 and 0.20 to 0.25, respectively. The wear volume of UHMWPE with radiation of 50 Mrad, 75 to 150 Mrad, and 200 Mrad was 70 to 80%, 18 to 25%, and 12 to 15%, respectively, in comparison to non-irradiated specimens. Elongation and tensile strength of UHMWPE with radiation of 100 Mrad decreased to 6%, and 50% of that without radiation, respectively. The hardness increased with increase of the radiation dose. From several kinds of tribological findings, mechanical strength tests, and studies of long-term clinical findings, it is concluded that approximately 200 Mrad is the optimum dose of gamma radiation for clinical use in total hip prostheses.
ISSN:0957-4530
1573-4838
DOI:10.1023/A:1018582027349