Energy resolution of an Omega-type monochromator and imaging properties of the MANDOLINE filter

We report on the energy resolution properties of an Omega-type monochromator. In a TEM/STEM setup with a MANDOLINE filter and extreme stability of the high voltage and the filter current, an energy resolution of 43 meV for 0.1 s exposure time and 87 meV for 100 s exposure time was measured at 200 kV...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Ultramicroscopy 2010-07, Vol.110 (8), p.971-980
Hauptverfasser: Essers, Erik, Benner, Gerd, Mandler, Thilo, Meyer, Stefan, Mittmann, Dieter, Schnell, Michael, Höschen, Rainer
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report on the energy resolution properties of an Omega-type monochromator. In a TEM/STEM setup with a MANDOLINE filter and extreme stability of the high voltage and the filter current, an energy resolution of 43 meV for 0.1 s exposure time and 87 meV for 100 s exposure time was measured at 200 kV with 40 meV monochromator slit width. The monochromized zero-loss peaks are additionally characterized by their edge steepness. Moreover a drop in the monochromized zero-loss peak by 10 3 after 260 meV can be obtained even without deconvolution. For small fields of view, the energy resolution mostly does not depend on the MANDOLINE filter. With the Corrected OMEGA filter an energy resolution of 41 meV was measured for 0.03 s exposure time at 200 kV with 30 meV monochromator slit width and 77 meV for 50 s exposure time at 80 kV with 40 meV monochromator slit width. Furthermore, the MANDOLINE filter’s setup and imaging properties are presented such as isochromaticity (
ISSN:0304-3991
1879-2723
DOI:10.1016/j.ultramic.2010.02.009