Liquid Chromatographic Analysis of Doxazosin in Human Serum with Manual and Robotic Sample Preparation
A specific method for the determination of the antihypertensive drug doxazosin in human serum is described. The method utilizes the related drug prazosin as an internal standard and is based on a simple extraction scheme followed by analysis by reversed-phase ion suppression high-performance liquid...
Gespeichert in:
Veröffentlicht in: | Journal of chromatographic science 1988-11, Vol.26 (11), p.570-573 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A specific method for the determination of the antihypertensive drug doxazosin in human serum is described. The method utilizes the related drug prazosin as an internal standard and is based on a simple extraction scheme followed by analysis by reversed-phase ion suppression high-performance liquid chromatography (HPLC) on an alumina-based column with fluorescence detection. The method is completely automated with a flexible robotic system for the analysis of drugs in biological fluids. The robotic automation of the method allows a 20% increase in the sample throughput and the savings of about 7 man-hours a day. Both the manual and robotic procedures yield precise quantitative results over the therapeutically relevant concentration range of 0.5 to 20 ng/mL of serum. |
---|---|
ISSN: | 0021-9665 1945-239X |
DOI: | 10.1093/chromsci/26.11.570 |