Effects of blood flow on laser probe temperature in human arteries
Laser recanalization using metal capped fibers occurs by thermal vaporization of occluding plaque. However, little is known about the effects of blood and flow on the temperature of the laser probe or the arterial wall during lasing. To study this, probe and arterial wall temperatures were measured...
Gespeichert in:
Veröffentlicht in: | Lasers in surgery and medicine 1988, Vol.8 (6), p.555-561 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Laser recanalization using metal capped fibers occurs by thermal vaporization of occluding plaque. However, little is known about the effects of blood and flow on the temperature of the laser probe or the arterial wall during lasing. To study this, probe and arterial wall temperatures were measured while a metal capped fiber, activated by an argon laser, was held stationary in a stenotic human peripheral artery. Arteries were perfused with saline and blood, and flow was varied from 0 to 140 cc/min. Probe temperatures were significantly higher in blood than in saline. However, the increased probe temperature achieved in blood was not transferred to the arterial wall. Increasing flow decreased probe temperature in both media, but again arterial wall temperatures were minimally affected. Thus, the presence of blood and flow may significantly affect heat generation and heat transfer during arterial recanalization using metal capped fibers. |
---|---|
ISSN: | 0196-8092 1096-9101 |
DOI: | 10.1002/lsm.1900080603 |