Assessment of thoracic gas volume by low-frequency ambient pressure changes in children

The validity of a new method for measuring thoracic gas volume (TGV) was studied in 69 children, 4-16 yrs old, including twelve normal children and 57 children with an obstructive (n = 38) or restrictive (n = 19) respiratory disease. The method consisted of applying very slow (0.05 Hz) sinusoidal va...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The European respiratory journal 1988-07, Vol.1 (7), p.594-599
Hauptverfasser: Peslin, R, Marchal, F, Gallina, C, Oswald, M, Crance, JP
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The validity of a new method for measuring thoracic gas volume (TGV) was studied in 69 children, 4-16 yrs old, including twelve normal children and 57 children with an obstructive (n = 38) or restrictive (n = 19) respiratory disease. The method consisted of applying very slow (0.05 Hz) sinusoidal variations of ambient pressure around the body (delta Pam = 40 cmH2O peak to peak) and studying the relationship between delta Pam and the resulting gas displacement at the mouth (Vaw): TGVapc = PB.delta Vaw/delta Pam.cos phi, where PB is barometric minus alveolar water vapour pressure and phi the phase angle between Pam and Vaw. Functional residual capacities derived from TGVapc (FRCapc) were compared to the values obtained by plethysmography (FRCplet) and by helium dilution (FRCdil). FRCapc did not differ significantly from FRCplet in either the entire group (1.75 +/- 0.62 l vs 1.79 +/- 0.45 l) or in the patient subgroups. However, with the new method a trend to slightly lower FRCs was seen in patients with the most obstruction (p less than 0.05). FRCdil was significantly lower than both FRCapc and FRCplet (p less than 0.001), particularly in children with obstruction. Significant correlations were found between the three methods (p less than 0.001). On the other hand, the method investigated requires that the subject breathe very regularly for a period of several minutes. This was rarely achieved, so that the reproducibility of the measurements was unacceptably low. At present, the method cannot be recommended for routine use in 4-16 yr old children.
ISSN:0903-1936
1399-3003
DOI:10.1183/09031936.93.01070594