The Downs' adjustable flow generator

Summary A venturi powered device for the generation of continuous positive airways pressure has been assessed. Some of its functional characteristics appear to be ideal, but the oxygen concentration and adequacy of flows require checking after each adjustment. A positive pressure relief valve and an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 1988-09, Vol.43 (9), p.766-769
1. Verfasser: FISHER, G. C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary A venturi powered device for the generation of continuous positive airways pressure has been assessed. Some of its functional characteristics appear to be ideal, but the oxygen concentration and adequacy of flows require checking after each adjustment. A positive pressure relief valve and an inlet valve are necessary for the safe use of the system and an airway pressure monitor is recommended.
ISSN:0003-2409
1365-2044
DOI:10.1111/j.1365-2044.1988.tb05752.x