Use of image enhancement during lithotripsy
Renal excursion during breathing is inevitable and is a cause of poor localization during extracorporeal shock-wave lithotripsy (SWL), which in theory might lead to poor treatment results. Eighty-one patients underwent lithotripsy treatment with and without the use of an image enhancement system des...
Gespeichert in:
Veröffentlicht in: | Journal of endourology 1996-08, Vol.10 (4), p.335-339 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Renal excursion during breathing is inevitable and is a cause of poor localization during extracorporeal shock-wave lithotripsy (SWL), which in theory might lead to poor treatment results. Eighty-one patients underwent lithotripsy treatment with and without the use of an image enhancement system designed for use with the Dornier MPL9000 lithotripter. This device contains a memory incorporated into a separate differential grayscale monitor, which allows the stone image to be stored. Shockwave release is enabled only when this image corresponds to the real-time image on the lithotripter ultrasound monitor. No improvement in success rates was found using this system, although upper-pole stones appeared to be fragmented more successfully. Overall, the results were favorable in both groups after a single treatment. Further work is needed to establish whether image enhancement is capable of improving the success rates and reducing the side effects of lithotripsy by better targeting. |
---|---|
ISSN: | 0892-7790 1557-900X |
DOI: | 10.1089/end.1996.10.335 |