Lead Explantation Late After Atrial Perforation
This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detect...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1996-08, Vol.19 (8), p.1268-1269 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This report describes the case of a patient in whom atrial perforation with penetration of the thoracic wall was diagnosed 2 years after the implantation of an Accufix lead. Despite this complication, atrial detection in the bipolar mode and ventricular pacing were normal. Digital fluoroscopy detected a fracture with extrusion of a short segment of the retention wire. The rupture of the retention wire might have been the result, but was not the cause of the perforation. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/j.1540-8159.1996.tb04200.x |