Rate Dependent Depression of Subsidiary Ventricular Impulse Formation-Cause of Stokes-Adams Attacks in a Patient with Rate Modulated Pacing

Following His bundle ablation and implantation of a rate adaptive pacemaker (Vitatron TX 911) a 52‐year‐old gentleman experienced several presyncopal attacks while driving a car. On examination pectoralis muscle contraction caused temporary pacemaker inhibition. Incremental overdrive stimulation dem...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1988-07, Vol.11 (7), p.1095-1097
Hauptverfasser: SCHMIDINGER, HERWIG, PROBST, PETER, WEBER, HEINZ, KALIMAN, JOSEF
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Following His bundle ablation and implantation of a rate adaptive pacemaker (Vitatron TX 911) a 52‐year‐old gentleman experienced several presyncopal attacks while driving a car. On examination pectoralis muscle contraction caused temporary pacemaker inhibition. Incremental overdrive stimulation demonstrated progressive depression idioventricular automaticity and was associated with similar symptoms following overdrive at high pacing rates. Following appropriate pacemaker programming the patient remained symptom free.