Measurement of X-ray image intensifier sharpness in the X-ray department
A method is presented for determining the modulation transfer function (MTF) of an image intensifier as it is found in the x-ray department. The image of an edge input into the image intensifier tube is photographed at the television camera port with a high quality camera and the photographic image...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 1988-01, Vol.33 (1), p.93-104 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A method is presented for determining the modulation transfer function (MTF) of an image intensifier as it is found in the x-ray department. The image of an edge input into the image intensifier tube is photographed at the television camera port with a high quality camera and the photographic image scanned with a microdensitometer. Densitometric conversion enables the determination of the edge spread function and further calculation produces the line spread function and the MTF. A correction is made for the camera/lens/film/developer response function. Results from the use of the technique on over 25 x-ray image intensifiers in New Zealand hospital x-ray departments are presented and discussed. Most of the image intensifiers had been in use for more than 7 years. The low-frequency drop values were similar to reported values with an average value of 22%, but modulation values at 1 cycle/mm (the average value was 29%) were generally lower than values previously reported, indicating that the sharpness performance in the clinical environment is inferior to claims for new or state-of-the-art intensifiers. |
---|---|
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/0031-9155/33/1/009 |