Correction to "Calorimetric Measurement of the Carbon Kerma Factor for 14.6-MeV Neutrons" by J. C. McDonald

The report corrected has a serious calculation error in the compensation for the γ-ray kerma of the mixed radiation field used to make the measurements. The error is explained and the value of the carbon kerma factor for 14.6-MeV neutrons is recalculated from the experimental data. Comparison of thi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation research 1988-02, Vol.113 (2), p.396-397
1. Verfasser: Goodman, Leon J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The report corrected has a serious calculation error in the compensation for the γ-ray kerma of the mixed radiation field used to make the measurements. The error is explained and the value of the carbon kerma factor for 14.6-MeV neutrons is recalculated from the experimental data. Comparison of this value to those determined by others suggests that the calorimetric technique described has a significant error.
ISSN:0033-7587
1938-5404
DOI:10.2307/3577213