Multicenter clinical validation of an on-line monitor of dialysis adequacy

Quantitation of hemodialysis by measuring changes in blood solute concentration requires careful timing when taking the postdialysis blood sample to avoid errors from postdialysis rebound and from recirculation of blood through the access device. It also requires complex mathematical interpretation...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American Society of Nephrology 1996-03, Vol.7 (3), p.464-471
Hauptverfasser: Depner, T A, Keshaviah, P R, Ebben, J P, Emerson, P F, Collins, A J, Jindal, K K, Nissenson, A R, Lazarus, J M, Pu, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Quantitation of hemodialysis by measuring changes in blood solute concentration requires careful timing when taking the postdialysis blood sample to avoid errors from postdialysis rebound and from recirculation of blood through the access device. It also requires complex mathematical interpretation to account for solute disequilibrium in the patient. To circumvent these problems, hemodialysis can be quantified and its adequacy assessed by direct measurement of the urea removed in the dialysate. Because total dialysate collection is impractical, an automated method was developed for measuring dialysate urea-nitrogen concentrations at frequent intervals during treatment. A multicenter clinical trial of the dialysate monitoring device, the Biostat 1000 (Baxter Healthcare Corporation, McGaw Park, IL) was conducted to validate the measurements of urea removed, the delivered dialysis dose (Kt/V), and net protein catabolism (PCR). The results were compared with a total dialysate collection in each patient. During 29 dialyses in 29 patients from three centers, the paired analysis of urea removed, as estimated by the dialysate monitor compared with the total dialysate collection, showed no significant difference (14.7 +/- 4.7 g versus 14.8 +/- 5.1 g). Similarly, measurements of Kt/V and PCR showed no significant difference (1.30 +/- 0.18 versus 1.28 +/- 0.19, respectively, for Kt/V and 42.3 +/- 15.7 g/day versus 52.2 +/- 17.4 g/day for PCR). When blood-side measurements during the same dialyses were analyzed with a single-compartment, variable-volume model of urea kinetics, Kt/V was consistently overestimated (1.49 +/- 0.29/dialysis, P < 0.001), most likely because of failure to consider urea disequilibrium. Because urea disequilibrium is difficult to quantitate during each treatment, dialysate measurements have obvious advantages. The dialysate monitor eliminated errors from dialysate bacterial contamination, simplified dialysate measurements, and proved to be a reliable method for quantifying and assuring dialysis adequacy.
ISSN:1046-6673
DOI:10.1681/ASN.V73464