Comparison of imaging accuracy at different MRI units based on phantom measurements

In this investigation standardized on-site phantom measurements at 27 MRI units were carried out. A scoring system was developed for rating both symmetry and distortion of the image. This study shows, with regard to the clinical use of MRI-assisted treatment planning in radiotherapy, that in more th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiotherapy and oncology 1995-12, Vol.37 (3), p.221-224
Hauptverfasser: Prott, F.J., Haverkamp, U., Willich, N., Resch, A., Stöber, U., Pötter, R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this investigation standardized on-site phantom measurements at 27 MRI units were carried out. A scoring system was developed for rating both symmetry and distortion of the image. This study shows, with regard to the clinical use of MRI-assisted treatment planning in radiotherapy, that in more than 85% of all planes (transversal, sagittal, coronal) the MRI units were equivalent to the standard known from literature.
ISSN:0167-8140
1879-0887