Comparison of imaging accuracy at different MRI units based on phantom measurements
In this investigation standardized on-site phantom measurements at 27 MRI units were carried out. A scoring system was developed for rating both symmetry and distortion of the image. This study shows, with regard to the clinical use of MRI-assisted treatment planning in radiotherapy, that in more th...
Gespeichert in:
Veröffentlicht in: | Radiotherapy and oncology 1995-12, Vol.37 (3), p.221-224 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this investigation standardized on-site phantom measurements at 27 MRI units were carried out. A scoring system was developed for rating both symmetry and distortion of the image. This study shows, with regard to the clinical use of MRI-assisted treatment planning in radiotherapy, that in more than 85% of all planes (transversal, sagittal, coronal) the MRI units were equivalent to the standard known from literature. |
---|---|
ISSN: | 0167-8140 1879-0887 |