Identifying oesophageal contents using electrical impedance tomography
Investigations have been carried out using the Sheffield mark II real-time EIT system, to look at changes in conductivity associated with swallowing. A ring of 16 electrodes was placed around the neck of 10 subjects, who then performed swallows with four liquids of different conductivities, ranging...
Gespeichert in:
Veröffentlicht in: | Physiological measurement 1995-11, Vol.16 (4), p.253-261 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Investigations have been carried out using the Sheffield mark II real-time EIT system, to look at changes in conductivity associated with swallowing. A ring of 16 electrodes was placed around the neck of 10 subjects, who then performed swallows with four liquids of different conductivities, ranging from water (sigma = 0.03 mS cm-1) to salty soup (sigma = 35.8 mS cm-1). Results showed that the conductive and non-conductive liquids could be distinguished. Bolus transit times were calculated from region of interest curves, and the average transit time for the 10 subjects was found to be 320 +/- 100 ms. |
---|---|
ISSN: | 0967-3334 1361-6579 |
DOI: | 10.1088/0967-3334/16/4/005 |