Optimized gradient waveforms for spiral scanning

Spiral scanning gradient waveforms can be optimized with respect to blurring from off‐resonance effects by minimizing the readout time. This is achieved by maximizing the gradient amplitude during the scan so that the edge of k‐space is reached as quickly as possible. Gradient hardware constraints a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 1995-08, Vol.34 (2), p.156-160
Hauptverfasser: King, Kevin F., K. F. Foo, Thomas, Crawford, Carl R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Spiral scanning gradient waveforms can be optimized with respect to blurring from off‐resonance effects by minimizing the readout time. This is achieved by maximizing the gradient amplitude during the scan so that the edge of k‐space is reached as quickly as possible. Gradient hardware constraints are incorporated by considering a circuit model for the gradient coil and amplifier. The optimized gradient waveforms are determined by a set of coupled differential equations. The resulting solutions have shorter readout time than solutions that do not consider the circuit model.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.1910340205