Minimizing work of breathing with continuous positive airway pressure and intermittent mandatory ventilation: An improved continuous low-flow system
Minimizing work of breathing (WOB) during intermittent mandatory ventilation (IMV) and continuous positive airway pressure (CPAP) is important as it facilitates weaning from mechanical ventilation. To minimize WOB, we devised a simple, continuous-flow CPAP-IMV system that uses a weighted, partially...
Gespeichert in:
Veröffentlicht in: | Critical care medicine 1987-07, Vol.15 (7), p.665-670 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Minimizing work of breathing (WOB) during intermittent mandatory ventilation (IMV) and continuous positive airway pressure (CPAP) is important as it facilitates weaning from mechanical ventilation. To minimize WOB, we devised a simple, continuous-flow CPAP-IMV system that uses a weighted, partially filled reservoir bag and operates efficiently at low fresh gas flow (FGF). We compared both the pattern and WOB of our system (FGF at 15 L/min) with a conventional continuous-flow CPAP/IMV system (FGF at 15 and 30 L/min) as well as with two relatively efficient demand-value systems, the Servo 900 B and 900 C. Six healthy male subjects were studied; tidal volumes (VT), flow, mouth pressure, and pleural pressure (Ppl) were measured. Ten breaths, matched for VT, from each subject on each system were selected for analysis. Mechanical WOB was estimated by integrating Ppl with respect to VT. The conventional continuous-flow system was associated with a high work/breath relative to the other systems (p < .001). The weighted reservoir system was associated with a significantly lower work/breath (p < .001), its performance approaching that of the Servo 900B. Work/breath was least with the Servo 900C (p |
---|---|
ISSN: | 0090-3493 1530-0293 |
DOI: | 10.1097/00003246-198707000-00008 |