MR imaging of large nonferromagnetic metallic implants at 1.5 T

To evaluate the feasibility of magnetic resonance (MR) imaging in patients with nonferromagnetic metallic implants, we imaged implants in vitro and in 15 patients. Image artifacts both in vitro and in vivo occurred only locally and consisted of image distortions, areas of total signal loss, and line...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of computer assisted tomography 1987-07, Vol.11 (4), p.678-683
Hauptverfasser: AUGUSTINY, N, VON SCHULTHESS, G. K, MEIER, D, BÖSIGER, P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To evaluate the feasibility of magnetic resonance (MR) imaging in patients with nonferromagnetic metallic implants, we imaged implants in vitro and in 15 patients. Image artifacts both in vitro and in vivo occurred only locally and consisted of image distortions, areas of total signal loss, and lines of high signal intensity. The artifacts were most prominent in areas where the implants exhibited edges or points, but overall image quality was good except for regions lying very close to the implants. The implants themselves appeared as structures without signal in all patients. We conclude that in patients with femoral head prostheses or osteosynthetic plates that are nonferromagnetic MR may be preferable to CT, where beam hardening artifacts usually degrade image quality severely in the entire field of view.