Combination sensor for measurement of light and temperature
This report describes light and temperature measurements made in a gel cell using a single probe. Glass pipettes were drawn to an outer tip diameter of 50 μm and a thermocouple was vapor deposited at the tip of the pipette. Direct absorption of light by the thermocouple produced near step changes in...
Gespeichert in:
Veröffentlicht in: | Lasers in surgery and medicine 1995, Vol.16 (4), p.397-400 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This report describes light and temperature measurements made in a gel cell using a single probe. Glass pipettes were drawn to an outer tip diameter of 50 μm and a thermocouple was vapor deposited at the tip of the pipette. Direct absorption of light by the thermocouple produced near step changes in temperature proportional to the fluence rate of the light. © 1995 Wiley‐Liss, Inc. |
---|---|
ISSN: | 0196-8092 1096-9101 |
DOI: | 10.1002/lsm.1900160412 |