Heating deep seated eccentrically located tumors with an annular phased array system: A comparative clinical study using two annular array operating configurations

Regional heating administered with an annular array to 12 patients with deep-seated advanced malignant disease eccentrically located in the lower abdomen and pelvis is compared based on the annular array operating configuration. One configuration (4 quadrants active) delivers radiofrequency power wi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of radiation oncology, biology, physics biology, physics, 1987, Vol.13 (1), p.83-94
Hauptverfasser: Samulski, Thaddeus V., Kapp, Daniel S., Fessenden, Peter, Lohrbach, Allen
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Regional heating administered with an annular array to 12 patients with deep-seated advanced malignant disease eccentrically located in the lower abdomen and pelvis is compared based on the annular array operating configuration. One configuration (4 quadrants active) delivers radiofrequency power with relative uniformity throughout the patient crossection. The other (2 quadrants active) allows the radiofrequency power deposition to be shifted preferentially into the eccentrically located treatment volume. Phantom measurements have been made to demonstrate the redistribution of radiofrequency power that results when the annular array is operated in these respective configurations. Systemic responses (i.e. oral temperature rise, changes in blood pressure, and heart rate) to these regional hyperthermia applications are compared and are not significantly different with respect to these heating configurations. Temperature data obtained during treatment sessions using these two annular array operating configurations are analyzed based on the fraction of measured tumor and normal tissue temperatures exceeding or equal to a given index temperature. Although the two quadrant configuration is more efficient in delivering power to the treatment volume, this analysis does not indicate a significant gain in therapeutic heating as a result of this preferential power deposition. Treatment tolerance and heterogeneity with respect to tissue type and blood flow remained the dominant limiting factors with regard to temperatures achieved.
ISSN:0360-3016
1879-355X
DOI:10.1016/0360-3016(87)90264-1