Retrieval of occluding unexpanded palmaz-schatz stent from a saphenous aorto-coronary vein graft

The Palmaz‐Schatz coronary stent is used frequently to reduce the rate of restenosis of balloon angioplasty in saphenous vein grafts. In many European centers, the stent, manually crimped on a balloon, is advanced across the stenosis without a protective sheath. This report describes a patient in wh...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Catheterization and cardiovascular diagnosis 1995-02, Vol.34 (2), p.159-161
Hauptverfasser: Rozenman, Yoseph, Burstein, Moshe, Hasin, Yonathan, Gotsman, Mervyn S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Palmaz‐Schatz coronary stent is used frequently to reduce the rate of restenosis of balloon angioplasty in saphenous vein grafts. In many European centers, the stent, manually crimped on a balloon, is advanced across the stenosis without a protective sheath. This report describes a patient in whom an attempt to deploy a stent in the orifice of a saphenous vein graft was complicated by dislodgement of the unexpanded stent from the balloon. The unexpanded stent caused immediate occlusion of the vein graft and severe ischemic symptoms. The stent was retrieved by inflation of another balloon in the graft, distal to the stent, pulling the balloon toward the guiding catheter, and then withdrawing the whole system.
ISSN:0098-6569
1097-0304
DOI:10.1002/ccd.1810340418