Simple method to measure total expiratory time constant based on the passive expiratory flow-volume curve
OBJECTIVEIn intubated, mechanically ventilated patients, inspiration is forced by externally applied positive pressure. In contrast, exhalation is passive and depends on the time constant of the total respiratory system. The expiratory time constant is thus an important determinant of mechanical ven...
Gespeichert in:
Veröffentlicht in: | Critical care medicine 1995-06, Vol.23 (6), p.1117-1122 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVEIn intubated, mechanically ventilated patients, inspiration is forced by externally applied positive pressure. In contrast, exhalation is passive and depends on the time constant of the total respiratory system. The expiratory time constant is thus an important determinant of mechanical ventilation. The aim of this study was to evaluate a simple method for measuring the expiratory time constant in ventilated subjects.
DESIGNProspective study using a lung simulator and ten dogs.
SETTINGUniversity hospital.
SUBJECTSCommercially available lung simulator and ten greyhound dogs.
INTERVENTIONSDifferent expiratory time constants were set on the lung simulator. In the dogs, the endotracheal tube was clamped to increase airways resistance by 22.5 cm H2 O/(L/sec) and the lungs were injured with hydrochloric acid to decrease total respiratory compliance by 16 mL/cm H2 O. This procedure resulted in a wide range of expiratory time constants.
MEASUREMENTS AND MAIN RESULTSPneumotachography was used to measure flow and volume. The ratio of exhaled volume and peak flow was calculated from these signals, corrected for the limited exhalation time yielding the "calculated expiratory time constant'' and compared with the actual expiratory time constant. The typical error was +/-0.19 sec for the lung simulator and +/-0.15 sec for the dogs.
CONCLUSIONSThe volume and peak flow corrected for limited exhalation time is a good estimate of the total expiratory time constant in passive subjects and may be useful for the titration of mechanical ventilation.(Crit Care Med 1995; 23:1117-1122) |
---|---|
ISSN: | 0090-3493 1530-0293 |