Design of the EPC-9, a computer-controlled patch-clamp amplifier. 2. Software
The EPC-9 computer-controlled amplifier has no front-panel controls; therefore the user interface to the EPC-9 patch-clamp amplifier is defined entirely by software. This paper describes various user interfaces that have been implemented, including a high-level programming interface, a user interfac...
Gespeichert in:
Veröffentlicht in: | Journal of neuroscience methods 1995, Vol.56 (2), p.203-215 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The EPC-9 computer-controlled amplifier has no front-panel controls; therefore the user interface to the EPC-9 patch-clamp amplifier is defined entirely by software. This paper describes various user interfaces that have been implemented, including a high-level programming interface, a user interface based on the PostScript language, and graphical user interfaces that control the EPC-9 from data-acquisition programs. Also described are the algorithms used for automatic adjustment of the C-Fast and C-Slow transient cancellation circuitry. An overview of the procedures that perform automatic testing and calibration is given. |
---|---|
ISSN: | 0165-0270 1872-678X |
DOI: | 10.1016/0165-0270(94)00129-5 |