Electrophilic 18F from a siemens 11 MeV proton-only cyclotron
Because more and more PET centres are using small proton cyclotrons there is a renewed interest in methods for the production of electrophilic 18F by proton irradiation of [ 18O]O 2. A method for the routine production of clinically useful quantities of [ 18F]F 2 having a specific activity of 35 Ci/...
Gespeichert in:
Veröffentlicht in: | Nuclear medicine and biology 1995, Vol.22 (1), p.111-116 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Because more and more PET centres are using small proton cyclotrons there is a renewed interest in methods for the production of electrophilic
18F by proton irradiation of [
18O]O
2. A method for the routine production of clinically useful quantities of [
18F]F
2 having a specific activity of 35 Ci/mmol has been developed and implemented using an 11 MeV proton cyclotron and [
18O]O
2. Based on the yield, purity, reproducibility, and specific activity of [
18F]F
2 this is the most efficient method reported thus far. |
---|---|
ISSN: | 0969-8051 1872-9614 |
DOI: | 10.1016/0969-8051(94)E0064-P |